DeFelsko PRBFNDS-D PosiTector 6000 Cabled Duplex Coating Systems Probe FNDS Only, Measures Coatings on Ferrous & Non-Ferrous Metals, 0-60 mils

See More DeFelsko PosiTector 6000

DeFelsko PRBFNDS-D PosiTector 6000 Cabled Duplex Coating Systems Probe FNDS Only, Measures Coatings on Ferrous & Non-Ferrous Metals, 0-60 mils

See More DeFelsko PosiTector 6000
Price:
$1,030.00
    warrenty

    Product Description
    DeFelsko PosiTector 6000 FNDS Cabled Spare Probe for Duplex Coatings For Use With 6000 Series
    Display Sold Separately

    The PosiTector 6000 consists of a body and a probe. A wide selection of interchangeable probes are available. All bodies accept all probes.

    The PosiTector 6000 FNDS probe measures the individual thicknesses of both the paint and the zinc layers in a duplex coating system with a single reading.

    Duplex coating systems use a combination of two corrosion protection systems – typically paint or powder coating over galvanized steel (hot-dip, electro or zinc spray metallizing). The resultant corrosion protection is superior to either protection system used independently.

    In Duplex mode, the PosiTector 6000 FNDS utilizes both magnetic (ferrous) and eddy current (non-ferrous) principles simultaneously to calculate and display the individual paint and zinc layer thicknesses. The magnetic principle is used to measure the combined paint/zinc thickness over the ferrous substrate, and the eddy current principle is used to measure the paint thickness over the non-ferrous zinc coating. The zinc thickness is calculated by subtracting the paint thickness from the combined paint/zinc thickness measurement.

    When disabled, the gage will operate like a conventional combination ferrous/non-ferrous instrument (similar to PosiTector 6000 FNS probe). Ideal for measuring non-magnetic coatings over steel and non-conductive coatings over non-ferrous metal substrates.

    Compatible with PosiTector Advanced (serial numbers 730,000+) and PosiTector Standard gage bodies with a color display. A Simple gage update may be required.


    line
    Specificationsline

    Range 0 - 60 mils
    0 - 1500 μ
    Minimum Zinc Thickness 0.5 mil (13 μ)
    Paint Layer Accuracy Mils ± (0.05 mils + 1%) 0 - 2 mils
    ± (0.05 mils + 1%) 0 - 2 mils
    Paint Layer Accuracy Microns ± (1 μ + 1%) 0 - 50 μ
    ± (2 μ + 1%) > 50 μ
    Zinc Layer Accuracy Mils ± (0.1 mils + 1%) 0.5 - 2 mils
    ± (0.2 mils + 1%) > 2 mils
    Zinc Layer Accuracy Microns ± (2 um + 1%) 13 - 50 um
    ± (4 um + 1%) > 50 um

    *Accuracies are stated as a fixed value plus a percentage of the gage's actual reading.


    line
    FAQ's - Frequently Asked Questions
    line

    Click to View DeFelsko PosiTector 6000 Frequently Asked Questions


    line
    Downloads
    line

    DeFelsko PosiTector 6000 Coating Thickness Gages - Full Instruction Manual
    DeFelsko PosiTector 6000 Coating Thickness Gages - Literature
    DeFelsko PosiTector 6000 Coating Thickness Gages - Quick Guide
    DeFelsko PosiTector 6000 Probe Details - Online
    DeFelsko PosiTector 6000 Coating Thickness Gages - Online Instructions Manual
    DeFelsko Accessories for Inspection Instruments - Online Catalog