Horiba B-751 LAQUAtwin Compact Calcium Ion Meter 3200456570

See More Horiba LAQUAtwin Compact Meters

Horiba B-751 LAQUAtwin Compact Calcium Ion Meter 3200456570

See More Horiba LAQUAtwin Compact Meters
This product has been discontinued

Product Description
This product has been replaced by the LaquaTwin Ca-11
The B-751 has been replaced by the Horiba LAQUAtwin Ca-11 Compact Water Quality Calcium Ion Meter 3200689161

Horiba B-751 LAQUAtwin Compact Calcium Ion Meter Part # 3200456570


Autocalibration function recognizes standards and is user-selectable for either one or two-point calibration. Measurement range 40ppm - 4000ppm. For submersible and special small sample or solid sample applications.

Temperature compensation

Only compact meter for a quick measurement of calcium ion using ion selective membrane.Flat sensor technology and 60 years of sensor engineering realizes a reliable and direct measurement of a drop of the sample from 0.3mL. Use sampling sheet (sold separately) to measure down to 0.05mL or more.Unique sensor mechanism realizes direct measurement of solid, viscous, powder samples. No need for a beaker to calibrate or measure. Just drop the standards and sample on the flat sensor, saves you time and minimize wasting your precious sample.This lightweight, pocket-sized meter is suitable for research and industrial applications in diverse fields.

Applications include:
Soil and sap testing for improved agriculture; food testing*

* Detection of ions in water only; electrolytes in oils cannot be measured. Ion measurement can be affected by non-target ions.

line
Features
line

  • Autocalibration function recognizes standards and is user-selectable for either one- or two-point calibrations.
  • Auto range change
  • Temperature compensation
  • IP67 waterproof
  • Auto hold
  • Automatic power off ?30 minutes)

line
Specifications
line

MODEL

B-751

Measurement principle

Ion electrode method

Minimum 
sample volume

0.3mL or more*1

Measurement range

40 to 4000 ppm(mg/L)(10-3 to 10-1 mol/L)

Display range

0 to 9900ppm(mg/L)*2

Range and Resolution
(Valid numbers)

?0 to 1.0ppm : 0.1ppm
?0 to 99ppm : 1ppm
?100 to 990ppm : 10ppm
?1000 to 9900ppm : 100ppm

Calibration

Two-point*3

Accuracy

±20% of reading value.*4

Functions

Auto range change?Temperature compensation?Waterproof*5?Reading locking ?Automatic power off ?30 minutes?

Display

Custom (monochrome) Digital LCD

Operating
temperature/humidity

5 to 40?C, 85% or less in relative humidity  (no condensation)

Power

CR2032 batteries (x2)

Battery life

Approx. 400 hours in continuous use

Main Material

ABS epoxy

Dimensions
/Mass

164mm × 29mm × 20mm  (excluding projections)?/ Approx. 50 g (meter only, without batteries)

Accessories included

2 CR2032 batteries/1 Pipette/Instruction manual/Quick manual/Storage case/Standard solution
(150ppm&2000ppm)(14mL)/5 pieces of Sampling sheetB

*1 Smaller amount(0.05mL or more) can be measured with the sampling sheet B.  (Please close the light shield cover. If a sample that contain particulate, please use ”Sampling sheet holder” (sold separately))
*2 When the measured value is out of the measurement range, the displayed value blinks. It should be used only as a guide.
*3 Selectable between one-point and two-point calibrations. 
*4 Repeatability in measurement of a standard solution after calibration using the same standard solution.
*5 IP67?no failure when immersed in water at a depth of 1 meter for 30 minutes. But the product can not be used underwater.


The ion electrode may be subject to interference from other ionic species. The selectivity coefficient reflects the affinity between each ionic species and the electrode, and is used to correct for the influence of competing ions. Please see the Interfering ion influence ? pH range for further details.

line
Related Products
line

Calcium Ion Sensor S050 related

line

Standard solution (Calcium Ion 2000ppm) Y051H related

line

Standard solution (Calcium Ion 150ppm) Y051L related

line

Sampling sheet B Y046 related

line

Sampling sheet holder (for LAQUAtwin) Y048 related

line
Downloads
line

B-751 LAQUAtwin Compact Calcium Ion Meter 3200456570 - Instruction Manual

B-751 LAQUAtwin Compact Calcium Ion Meter 3200456570 - Concept Book

B-751 LAQUAtwin Compact Calcium Ion Meter 3200456570 - Catalog

B-751 LAQUAtwin Compact Calcium Ion Meter 3200456570 - Quick Manual

B-751 LAQUAtwin Compact Calcium Ion Meter 3200456570 - Water Quality

B-751 LAQUAtwin Compact Calcium Ion Meter 3200456570 - Data Sheet

line