Horiba 6583-10C Calcium ion electrode (combination)

See More Horiba LAQUA Ion Electrodes

Horiba 6583-10C Calcium ion electrode (combination)

See More Horiba LAQUA Ion Electrodes
This product has been discontinued

Product Description
This product has been replaced by the Laqua 6583S-10C
The 6583-10C has been replaced by the Horiba Laqua 6583S-10C Combination Calcium Ion Electrode, 3200697410
The response time is the time which is required to reach 90%

6583-10C Calcium ion electrode (combination)

line
Specifications
line

Measuring range

0.4 to 40,080 mg/L Ca2+
(10-5 to 1 mol/L) Ca2+

pH range

4.0 mg/L (10-4 mol/L) Ca2+
pH 5 to 11

Applicable temperature range

0 to 50?

Response time (90%)

Within 15 seconds*4

Applicable reference electrode

-

Selection coefficient

Fe3+ = 0.1 Fe2+?Zn2+ = 1 
Sr2+ = 50 Ni2+?Cu2+ = 70 Co2+ = 350
Mn2+ = 500 Mg2+ = 1,000
Na+?K+?Ba2+?NH4+ = over 1,000

  • The response time is the time which is required to reach 90% response when the ion concentration is gradually changed from 10-4 mol/L to 10-2 mol/L with the solution stirred.

Exception:

*1: 90% response when ion concentration is changed to 10-6 mol/L ? 10-2 mol/L
*2: 95% response when ion concentration is changed to 10-3 mol/L ? 10-1 mol/L
*3: 95% response when ion concentration is changed to 10-4 mol/L ? 10-2 mol/L
*4: 95% response when ion concentration is changed to 10-4 mol/L ? 10-1 mol/L

  • The selection coefficient is a ratio of the limit concentration of coexisting ions (mol/L) to the ion concentration to be measured (mol/L); The value of 1000 means that the coexisting ions can be permitted up to 1000 times the ion measured and ''not acceptable'' means that chemical change occurs in the solid response membrane.