Horiba Laqua 8005-10C Bromide Ion Electrode, # 3014094397

See More Horiba LAQUA Ion Electrodes

Horiba Laqua 8005-10C Bromide Ion Electrode, # 3014094397

See More Horiba LAQUA Ion Electrodes
This product has been discontinued

Product Description
Horiba Laqua 8005-10C Bromide Ion Electrode, # 3014094397
Measuring range: 0.8 to 80,000 mg/L Br- | (10-5 to 1 mol/L) Br-
pH Range: 800 mg/L (10-2 mol/L) Br- | pH 1.5 to 11.5
Applicable temperature range: 0 to 50℃

The Horiba Laqua 8005-10C Bromide Ion Electrode, # 3014094397 is an Ion Selective Electrode - the reference electrode and ion sensing electrode are built in one body.

HORIBA's superior electrode technology has produced unparalleled tough pH glass bulbs and unique flat sensors. Electrodes have different designs to cater a wide range of applications - from pure water to complex samples.


line
Specifications
line

Measuring range

0.8 to 80,000 mg/L Br-
(10-5 to 1 mol/L) Br-

pH range

800 mg/L (10-2 mol/L) Br-
pH 1.5 to 11.5

Applicable temperature range

0 to 50℃

Response time (90%)

Within 5 seconds

Applicable reference electrode

2565A

Selection coefficient

S2O32-、l-、S2-、CN- = Not acceptable
MnO4- = 1
Cl-、PO42- = 100
F-、NO3-、SO42- = 1,000

  • The response time is the time which is required to reach 90% response when the ion concentration is gradually changed from 10-4 mol/L to 10-2 mol/L with the solution stirred. Exception:
    • *1: 90% response when ion concentration is changed to 10-6 mol/L ∼ 10-2 mol/L
    • *2: 95% response when ion concentration is changed to 10-3 mol/L ∼ 10-1 mol/L
    • *3: 95% response when ion concentration is changed to 10-4 mol/L ∼ 10-2 mol/L
    • *4: 95% response when ion concentration is changed to 10-4 mol/L ∼ 10-1 mol/L
  • The selection coefficient is a ratio of the limit concentration of coexisting ions (mol/L) to the ion concentration to be measured (mol/L); The value of 1000 means that the coexisting ions can be permitted up to 1000 times the ion measured and ''not acceptable'' means that chemical change occurs in the solid response membrane.

  • line
    Downloads
    line

    Horiba Laqua 8005-10C Bromide Ion Electrode, 3014094397 - Brochure

    ADD TO CART THE ITEMS YOU WOULD LIKE TO PURCHASE
    Models Price
    Discontinued